• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. VAS HD AC 4MM X 30 DEG; NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. VAS HD AC 4MM X 30 DEG; NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) Back to Search Results
Model Number 72202959
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Foreign zip code: (b)(6).Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that the scope had the distal lens cracked.No case involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection found the outer surface dented, distal tip and fiber damage, and a cracked distal negative lens.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat event.The complaint was confirmed.Factors that could have contributed to the reported event include an impact event inconsistent with normal use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAS HD AC 4MM X 30 DEG
Common Device Name
NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
MDR Report Key10869915
Report Number3003604053-2020-00151
Device Sequence Number3049010
Product Code EOB
UDI-Device Identifier00885554018783
UDI-Public00885554018783
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K043395
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72202959
Device Catalogue Number72202959
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/27/2020
Supplement Date Received by Manufacturer04/16/2021
Initial Report FDA Received Date11/19/2020
Supplement Report FDA Received Date04/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-