• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. RENASYS GO CANISTER 300ML; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. RENASYS GO CANISTER 300ML; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Catalog Number 66800914
Device Problem Obstruction of Flow (2423)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/28/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that canister is blocking the negative pressure coming from the device to the wound bed, the exudate is not passing from the wound bed to the canister.After a canister of the same batch had the same issue, a different canister batch was tried and no issue.This issue was identified during procedure and the procedure was completed with a canister of a different batch.The canister from the hospital was tested by s+n team with a different renasys go device, to test if the issue was with the device and not the canister, but the same issue of canister being blocked was seen.No harm or injury reported.
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.All provided information has been reviewed and we have not been able to establish a relationship between the device and the reported event or determine a root cause.Probable causes may be due to the reported blockage or a component failure.The ifu offers further guidance.The manufacturing records show no evidence that the product did not meet specification at the time of manufacture.A complaint history review found further instances of the reported event.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS GO CANISTER 300ML
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10871214
MDR Text Key217289281
Report Number8043484-2020-03942
Device Sequence Number1
Product Code OMP
UDI-Device Identifier40565125695
UDI-Public40565125695
Combination Product (y/n)N
PMA/PMN Number
K152163
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66800914
Device Lot NumberD1911919
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/28/2020
Initial Date FDA Received11/19/2020
Supplement Dates Manufacturer Received07/01/2021
Supplement Dates FDA Received07/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-