| Model Number |
72200731 |
| Medical Device Problem Code |
Flaked (1246)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopic surgery, when using the acromioblaster burr on full forward, an extremely loud high pitch screeching sound appeared.Like the metal was grinding.Metal fragments were present, although the malfunction occurred outside the patient.The procedure was completed without significant delay using a back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.Microscopic inspection of the burr revealed worn cutting edges and flakes in the cannula.A functional evaluation revealed a loud high pitch noise when the burr was operated in forward and reverse.The complaint was confirmed.Factors which could have contributed to the reported event include contact with a metallic object, excessive side-loading, or operating the device without proper irrigation.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.
|
| |
|
Search Alerts/Recalls
|