MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVCE REP DYONICSPOWERMINI HAND CONTROLS; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVCE REP DYONICSPOWERMINI HAND CONTROLS; SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 72201500S
Device Problem Protective Measures Problem (3015)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Event Description
It was reported that the dyonics powermini displayed "unknown handpiece" error message.No case involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, intended for use in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A complaint history review concluded this was a repeat issue.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Manufacturer Narrative
Updated, device received the reported device, intended for use in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of unknown handpiece error could not be reproduced.Product passed functional and high speed testing (100-10,000 rpms) for 5 minutes each in forward, reverse and oscillate directions.Unit passed functional tests on the dii and dii eip test control units with footswitch.Unknown handpiece error did not occur during functional testing.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REP DYONICSPOWERMINI HAND CONTROLS
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10871484
MDR Text Key217719003
Report Number1643264-2020-01885
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010654823
UDI-Public03596010654823
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 02/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72201500S
Device Catalogue Number72201500S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2020
Date Manufacturer Received02/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-