| Model Number |
10706 |
| Medical Device Problem Codes |
Device Displays Incorrect Message (2591); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of an angiojet spiroflex catheter.The pump assembly, effluent/supply line, shaft, tip and spike line were visually examined for damage or any irregularities.It was noticed when returned there was 25cc of blood in the waste bag and the device.The shaft showed multiple kinks.Functional testing was competed.The device was inserted into the test ultra drive unit console and the device was tested.The device would not prime due to the damaged shaft.There were 2 leaks along the shaft approximately 110cm and 114cm from the tip.Shaft damage at these locations in the outer shaft was noticed.The pressure was verified and was below the normal operating range.No other failures or error codes were noticed during the testing.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that an error message was displayed via medwatch (b)(4).An angiojet spiroflex catheter was used for a thrombectomy procedure.During the procedure, it was noted that "supply line failure" error message occurred.The catheter was re-primed and a kinked was noticed in the tubing and the saline line was unable to kept straightened.The procedure was completed with another of the same device.No patient complications were reported and patient was fine.
|
| |
|