NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number MS1-4570S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Wound Dehiscence (1154); Fever (1858)
Date of Event 10/14/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not been returned for evaluation.Root cause is unable to be determined at this time.Device has not been returned.
 
Event or Problem Description
Information was received that a revision procedure was performed on (b)(6) 2020.As per the reporter the patient was febrile post-op and was found to have purulence thoracic wound dehiscence found at the expansion end of the rod.The patient was treated with antibiotics and had a culture and immunodeficiency work up.
 
Additional Manufacturer Narrative
Evaluation summary: no device problems were identified.The returned rod was observed to be partially distracted.No contamination or scratches were visible.The rod was functionally tested and met specifications.During the functional testing no fluid/material was extruded out of the actuator.Review of the device history records revealed no discrepancies related to this complaint; the rods were manufactured in accordance with the specified requirements and met all release criteria.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
MDR Report Key10872388
Report Number3006179046-2020-00473
Device Sequence Number16867668
Product Code HGB
UDI-Device Identifier00812258026332
UDI-Public812258026332
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K171791
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberMS1-4570S
Device Lot Number0071501-AAC
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/19/2020
Supplement Date Received by Manufacturer03/15/2021
Initial Report FDA Received Date11/19/2020
Supplement Report FDA Received Date03/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age6 YR
Patient Weight15
-
-