| Model Number |
000724 |
| Medical Device Problem Codes |
Break (1069); Fluid/Blood Leak (1250); Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 06/2023).
|
| |
|
Event or Problem Description
|
|
It was reported that approximately thirty-five days post gastrostomy tube placement, the tube allegedly dislodged and fell out of the patient.It was revealed that the balloon ruptured.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately thirty-five days post gastrostomy tube placement, the tube allegedly dislodged and fell out of the patient.It was revealed that the balloon ruptured.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a complaint history review was performed.This is the seventh complaint reported for this product/lot number combination.The device history records were reviewed and there was nothing found to indicate there was a manufacturing related cause for these reports.Investigation summary: the device was not returned for evaluation.The investigation is inconclusive for the reported tube dislodged, rupture and break issues.A definitive root cause could not be determined.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|