| Model Number |
33302 |
| Medical Device Problem Code |
Nonstandard Device (1420)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the 6 male external catheters had too much adhesive and were unrolling.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed.Visual evaluation noted 2 unopened spirit mecs were received.No obvious defects were noted.Further testing required.Adhesive peel test was performed.Samples were cut to the required width (0.70" +/- 0.05").Adhesive peel strength was found to be within specification (0.60-2.10lbf).One of the samples tested did not generate a reading, and therefore only one reading was able to be generated.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.Labeling review was not performed because labeling could not have prevented the reported failure.Correction: h.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The actual/suspected device was inspected.
|
| |
|
Event or Problem Description
|
|
It was reported that the 6 male external catheters had too much adhesive and were unrolling.
|
| |
|