Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Visual analysis of the returned device identified: deformation and creases.A weight test of the device was verified and the device was within specification.Cloudy after autoclave disinfection process in the gel.Based on the device analysis the final assessment is: no issues found related with the manufacturing.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii and patient's concern with the product.
|