| Model Number |
190-31-03 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/11/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pending evaluation.Concomitant medical device: (cn: 142-32-00, sn: (b)(4)) cocr femoral head 32mm +0 offset 12/14.(cn: 01-030-40-0232, sn: (b)(4)) alteon xle liner natural g2 32mm.
|
| |
|
Event or Problem Description
|
|
As reported, approximately i year postop the initial tha, this female patient was revised due to a loose stem.The cocr head was removed after dislocating, the stem was removed, the poly liner was removed to check the cup, a new poly was implanted, and a monobloc stem was re-implanted along with a new head.The patient received a new liner, a 16x245 monobloc, with a new head.The patient appeared to have prior osteoarthritis.Patient was last known to be in stable condition following the event.Devices to be returning.¿.
|
| |
|
Additional Manufacturer Narrative
|
|
The revision reported was likely the result of an insufficient bond between the implant and the bone which led to aseptic (non-infected) loosening of the femoral stem.The most probable root cause associated with the reported event of "loosening - femoral stem¿ is associated with weakened integration of the device at the bone-implant interface due to loss of fibrous and/or bony tissue and leading to compromised anchorage of the device.
|
| |
|