• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO BONE PIN 4.0MM X 127MM; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO BONE PIN 4.0MM X 127MM; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number PFSD01008
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Codes No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a navio demo, the pin was bent.No patient was involved.
 
Additional Manufacturer Narrative
H6: the navio bone pin, p/n pfsd01008, lot tu60072-01, intended for use in treatment, was returned for evaluation.A relationship between the reported event and the device was established.The reported issue was visually confirmed.The bone pin is bent approximated to a 45 degree angle in the the start of the threads.A review of manufacturing records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.The most likely cause of this event was external forces and/or pressure apply to the pin.Care and caution should be exercised during the surgical site setup, surgery and tear down to protect the device from an unforeseen force, such as falling onto a hard surface or damage.Refer to the navio surgical system user¿s manual for proper handling.Components break / wear or spinning bur contacts inside of guard causing flaking / material dust.Per complaint, the component is a demo one thus no patient harm or injury was report.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO BONE PIN 4.0MM X 127MM
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
7634524980
MDR Report Key10874821
Report Number3010266064-2020-02013
Device Sequence Number3315677
Product Code OLO
UDI-Device Identifier00885556628737
UDI-Public00885556628737
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K191223
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPFSD01008
Device Catalogue NumberPFSD01008
Device Lot NumberTU60072-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2020
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2020
Supplement Date Received by Manufacturer11/15/2021
Initial Report FDA Received Date11/20/2020
Supplement Report FDA Received Date11/16/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-