| Model Number |
806455 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/26/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Per the manufacturer's subsidiary, the perfusionist found that the reservoir volume was increased even when the occluder was fully closed.It looked like the plunger was not coming out completely.This complaint is related to (b)(4) / mfr#1828100-2020-00424.
|
| |
|
Event or Problem Description
|
|
It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, the occluder did not fully occlude.As mitigation, manual occlusion was done using forceps.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
| |
|
Event or Problem Description
|
|
Additional information was received from the manufacturer's clinical specialist that the issue occurred during priming of the device.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: 81 evaluation is in progress, but not yet concluded.
|
| |
|
Additional Manufacturer Narrative
|
|
During laboratory analysis, the product surveillance technician (pst) did not observe any issues.A functional test was performed on the occluder and observed no water level change in the tubing over time.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported complaint was confirmed based on a video that was sent in by the user but was unable to be duplicated during testing.The service repair technician (srt) could not duplicate the complaint.Testing was performed and the srt observed the occluder head to function as intended.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
| |
|