• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. FOUNDATION KNEE; FEMUR, POSTERIOR STABILIZED, FOUNDATION, SIZE3, RT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. FOUNDATION KNEE; FEMUR, POSTERIOR STABILIZED, FOUNDATION, SIZE3, RT Back to Search Results
Model Number 214-01-003
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 10/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery - due to unknown reason.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as unknown.The previous surgery and the surgery detailed in this event occurred 4 years and 9 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to unknown reason.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, patient bone deterioration, inadequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Additional Manufacturer Narrative
Corrected data: see g1 - new contact for encore medical l.P.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION KNEE
Common Device Name
FEMUR, POSTERIOR STABILIZED, FOUNDATION, SIZE3, RT
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key10875544
Report Number1644408-2020-01062
Device Sequence Number4329399
Product Code JWH
UDI-Device Identifier00888912110532
UDI-Public(01)00888912110532
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K964008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2017
Device Model Number214-01-003
Device Catalogue Number214-01-003
Device Lot Number099H1001
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/22/2020
Supplement Date Received by Manufacturer01/20/2021
01/20/2021
Initial Report FDA Received Date11/20/2020
Supplement Report FDA Received Date02/17/2021
03/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
360-15-502 LOT 361B1020; 360-15-502 LOT 361B1020
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
-
-