| Model Number |
M00562693 |
| Medical Device Problem Codes |
Break (1069); Entrapment of Device (1212); Retraction Problem (1536)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/04/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a sensation short throw was used during a colonoscopy procedure performed on (b)(6) 2020 according to the complainant, during the procedure, the snare handle mechanism broke and the snare failed to close properly during the polypectomy.Reportedly, the snare loop became stuck in the polyp and the tech was forced to cut the snare wire by the handle in order to force pull the loop into the catheter.The snare was successfully removed from the patient using suction.The procedure was completed with a different captivator snare.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be ok.
|
| |
|