LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Pumping Stopped (1503)
Patient Problem No Patient Involvement (2645)
Event Date 10/23/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement. The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402). Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). Field service technician dispatched to the facility confirmed the issue and solved it by patient pump replacement. Subsequent functional verification testing was completed without further issues and the unit was returned to service. The issue was most likely caused by environmental conditions in which the pumps operate. Water infiltration or water formation due to condensation or high temperatures and high level of humidity in the stationary phase may led to bearing damage not allowing the pump the rotate freely.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t had a faulty pump on patient circuit 1. The issue was identified during the routine cleaning. There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM 80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM 80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key10875818
MDR Text Key217437461
Report Number9611109-2020-00645
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeMY
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 11/16/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/20/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number16-02-80
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received10/23/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/14/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-