| Model Number |
100/870/080 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/01/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot number is unknown at this time.
|
| |
|
Event or Problem Description
|
|
Information was receive indicating that the portex tubes blue line ultra worked properly during the pre-use check, but during the use of the product the cuff did not inflate and had a pinhole in it.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: returned device was received in used condition, decontaminated, without its original packaging.During the evaluation of the device per visual inspection it is confirmed the cuff has a pinhole.The customer reported condition was confirmed.The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.
|
| |
|