Model Number 100/870/080
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The lot number is unknown at this time.
 
Event or Problem Description
Information was receive indicating that the portex tubes blue line ultra worked properly during the pre-use check, but during the use of the product the cuff did not inflate and had a pinhole in it.There was no patient injury.
 
Additional Manufacturer Narrative
Other, other text: returned device was received in used condition, decontaminated, without its original packaging.During the evaluation of the device per visual inspection it is confirmed the cuff has a pinhole.The customer reported condition was confirmed.The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
MDR Report Key10876523
Report Number3012307300-2020-11633
Device Sequence Number14177023
Product Code BTO
UDI-Device Identifier15019315038808
UDI-Public15019315038808
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K030570
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number100/870/080
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/21/2020
Supplement Date Received by Manufacturer11/28/2020
Initial Report FDA Received Date11/20/2020
Supplement Report FDA Received Date12/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-