| Model Number |
1011487-15 |
| Medical Device Problem Codes |
Failure to Advance (2524); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a 90% stenosed, heavily tortuous de novo lesion in the left vertebral artery.The 4.0x15 mm rx herculink elite stent delivery system (sds) failed to cross due to the anatomy.The sds was removed without any issues.Once outside the patient anatomy the stent dislodged.Another 4.0x15mm rx herculink elite sds was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual analysis was performed on the returned product.The reported failure to advance was not tested as it was based on case circumstances.The stent dislodgment was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints.The investigation determined that the reported failure to advance and stent dislodgment were due to case circumstances.It is likely that anatomical conditions contributed to the reported failure to cross.It is likely that the stent became compromised (loosened) on the balloon during the failed attempt to advance resulting in dislodgment of the stent once removed from the anatomy.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.Na.
|
| |
|
Search Alerts/Recalls
|