C.R. BARD, INC. (COVINGTON) -1018233 BARD TOUCHLESS PLUS UNISEX PRE-LUBRICATED RUBBER URETHRAL CATHETER KIT Back to Search Results
Model Number 4A5044
Medical Device Problem Code Improper Chemical Reaction (2952)
Health Effect - Clinical Code Pain (1994)
Date of Event 10/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the intermittent catheters in the kits were very dry and hence patient had hard time with insertion.Stated that the patient felt pinching from insertion.Patient had 10 used catheters from the same lot that had caused issues.No medical intervention was reported.
 
Event or Problem Description
It was reported that the intermittent catheters in the kit were very dry and hence the patient had a hard time during the insertion.It was stated that the patient felt pain during the insertion.The patient had 10 used catheters from the same lot that had caused the issues.No medical intervention was reported.
 
Additional Manufacturer Narrative
The reported event was unconfirmed as the problem could not be reproduced.Visual evaluation of the returned sample noted nine unopened (within original package) unused red rubber touchless catheters were received.Visual inspection of the sample noted no obvious visible defects.A large amount of lubricant was noted at the tip and inside the bag prior to dispensing the catheter.After pushing the catheter out of the bag which coated the catheter in lubricant the catheters were wiped 10 times down the shaft of the catheter to judge its lubricity.After 10 wipes all nine catheters were still lubricious.A potential root cause for this failure mode was unable to determine because the reported event was unconfirmed.The device had met relevant specifications.The returned products were not used for the treatment.The product did not caused the reported failure.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "instructions for use: preparing for catheterization: 1.Open kit and remove contents.2.The gloves are not necessary to maintain aseptic conditions during the procedure.The gloves are for the operators protection.Male/female catheterization: 1.Remove cap from insertion tip while squeezing tabs.Save cap for closing the bag.2.Slide the catheter halfway into the insertion tip.3.Male - prepare the male urethral meatus and the surrounding area with povidone-iodine swabs provided.Female - prepare the female urethral meatus by holding the outer labia apart and prep the urethral meatus and surrounding area with povidone iodine swabs provided.4.Male - holding the penis, advance the insertion tip into the urethra no further than the flange base.Female - spread the inner labia, advance the insertion tip into the urethra no further than the flange base.Release the inner labia.5.Place your nondominant hand on the finger control guide to stabilize catheter in urethra.With your dominant hand, grasp catheter through bag approximately 1 inch below the finger control guide and push catheter into urethra.The catheter should be introduced by short, repetitive pushing motions.Repeat motions until catheter reaches bladder and urine starts to flow.6.Allow urine to flow freely, making certain the catheter guide is elevated at least 4 inches above lower portion of the bag.Allow flow until bladder is empty or until bag is filled.Precaution: it is recommended that the collection bag be held.The catheter could possibly separate from the collection bag when urine increases weight of the bag.7.Withdraw the catheter from the urethra.Remove the remaining portion of the catheter from the collection bag by pulling it through the insertion tip.Note: the catheter is designed to pass through the insertion tip.8.The filled collection bag may be closed by replacing the cap over the insertion tip.Make sure the cap snaps on securely.Specimen collection: to collect a specimen, obtain a sample cup or tube and alcohol wipe.Remove guide tip by pulling tab(s) upward from bag.Wipe port with alcohol.Pour specimen through port at top of bag into cup/tube.Draining bag: remove insertion tip by pulling tab(s) upward from bag.Drain urine through port at top of bag." h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD TOUCHLESS PLUS UNISEX PRE-LUBRICATED RUBBER URETHRAL CATHETER KIT
Common Device Name
URETHRAL CATHETER KIT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key10877782
Report Number1018233-2020-21131
Device Sequence Number3803337
Product Code FCM
UDI-Device Identifier00801741030512
UDI-Public(01)00801741030512
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K910653
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/31/2022
Device Model Number4A5044
Device Catalogue Number4A5044
Device Lot NumberNGEQ2242
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2021
Initial Date Received by Manufacturer 10/28/2020
Supplement Date Received by Manufacturer04/06/2021
Initial Report FDA Received Date11/20/2020
Supplement Report FDA Received Date04/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-