BIOTRONIK SE & CO. KG SENTUS PROMRI OTW QP L-75; LV LEAD Back to Search Results
Model Number 401182
Medical Device Problem Codes Failure to Capture (1081); High impedance (1291); Ambient Noise Problem (2877)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/12/2020
Type of Reportable Event Malfunction
Event or Problem Description
Lv lead did not capture, impedance was higher than 3000 ohms and noise was present on iegm channel.The patient was hospitalized and the lv lead replaced.Both the lv lead and device have been substituted.
 
Additional Manufacturer Narrative
Upon receipt, the lead under complaint was subjected to an extensive analysis.The performance of the device was scrutinized, including a visual and electrical inspection.The visual inspection revealed 32 cm distal to the is4 connector pin that the lead body was found squeezed and deformed.In this area the conductor coils leading to the ring electrodes and to the lead tip were found fractured, which is assumed to be the root cause of the reported clinical observation.Based on the characteristics, as well as the location of the damage, it is reasonable to assume that the lead was subject to excessive mechanical forces as the result of clavicular first rib entrapment.The manufacturing process for this device was re-investigated.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing process.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTUS PROMRI OTW QP L-75
Common Device Name
LV LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
MDR Report Key10878085
Report Number1028232-2020-05073
Device Sequence Number11844567
Product Code OJX
UDI-Device Identifier04035479138431
UDI-Public04035479138431
Combination Product (Y/N)N
PMA/510(K) Number
P070008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2019
Device Model Number401182
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2020
Initial Date Received by Manufacturer 11/19/2020
Supplement Date Received by Manufacturer03/01/2021
Initial Report FDA Received Date11/20/2020
Supplement Report FDA Received Date03/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-