Model Number 9040 B
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/19/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during the pre-operation, the surgeon tested the shunt, and it had water flow.When the shunt was used on the patient during the operation, there was no water flowing through.The surgeon opened a new shunt set, and the operation was carried on without any injuries.This resulted in a 10 minute delay in the procedure.The patient's status at the time of the report was alive-no injury.
 
Additional Manufacturer Narrative
The returned device was patent.The valve passed requirements for reflux, leakage and pressure flow/preimplantation testing.No signs of damage or debris were observed on the valve.The laboratory personnel could not replicate the failure condition of obstruction as observed in the field.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CSF-FLOW CONTROL SHUNT KIT, BURR HOLE 12 MM, MEDIUM PRESSURE
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
5290 california ave
irvine CA 92617
MDR Report Key10878289
Report Number2021898-2020-00343
Device Sequence Number17094589
Product Code JXG
UDI-Device Identifier00643169466357
UDI-Public00643169466357
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K831678
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2024
Device Model Number9040 B
Device Catalogue Number9040 B
Device Lot NumberE59641
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2021
Initial Date Received by Manufacturer 11/19/2020
Supplement Date Received by Manufacturer02/12/2021
Initial Report FDA Received Date11/20/2020
Supplement Report FDA Received Date02/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age13 MO
-
-