• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-ASD-006
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/03/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Further information regarding this event has been requested.If returned, investigation results will be provided in a subsequent submission.
 
Event or Problem Description
On (b)(6) 2020, a 6mm amplatzer septal occluder (aso) was implanted in the patient's defect.On (b)(6) 2020, there was no symptom on the patient, but a routine ultrasound cardiography was obtained.Then the aso was found to have become dislodged from the defect.By abdominal ct, the aso was located near the common iliac artery.By using a 9fr sheath and a double snare technique, the aso was removed from the patient percutaneously without much bleeding of the patient.The patient's general condition is stable, and the patient will be closely monitored.The patient is stable.
 
Additional Manufacturer Narrative
The reported event of device migration could not be confirmed.A more comprehensive assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
MDR Report Key10879570
Report Number2135147-2020-00496
Device Sequence Number2863113
Product Code MLV
UDI-Device Identifier00811806010038
UDI-Public00811806010038
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Model Number9-ASD-006
Device Catalogue Number9-ASD-006
Device Lot Number7028366
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/03/2020
Supplement Date Received by Manufacturer12/18/2020
Initial Report FDA Received Date11/20/2020
Supplement Report FDA Received Date12/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age11 YR
-
-