DEPUY SYNTHES PRODUCTS LLC SMALL BATTERY DRIVE II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Model Number 532.110 |
Device Problem
Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/05/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.Udi: (b)(4).
|
|
Event Description
|
It was reported that during a veterinary surgical procedure, it was observed that the small battery drive device was making a rattling noise.It was reported that the user was unable to proceed and the surgery was completed at a later date.It was unknown if a spare device was available for use.There was no human patient involvement as this was a veterinary procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was not reported, however it was reported that the event occurred in (b)(6) 2020.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Corrected data: the date returned for evaluation was reported as nov 10, 2020, the correct date is nov 11, 2020.
|
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: this device was returned for evaluation; however, during pre-repair assessment, it was determined that the device passed all operational specifications and no failure was identified.Therefore, the reported condition was not confirmed and an assignable root cause was not determined.
|
|
Search Alerts/Recalls
|
|
|