• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MICRORAPTOR RGNSB SA UB II #1 STR BL CB; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MICRORAPTOR RGNSB SA UB II #1 STR BL CB; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72204984
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Information (3190)
Date of Event 10/27/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a shoulder arthroscopy stabilization, the depth of the "microraptor drill (1,6mm)" in bone when crossing the cartilage was not deep enough to anchor well the "microraptor regenesorb (sa, ub ii, #1 str bl cb)".Therefore, when knotting the anchor got released easily and had to be removed from the patient.It is unknown if a back-up device was available or if there was a delay in the surgical procedure.No additional complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRORAPTOR RGNSB SA UB II #1 STR BL CB
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key10879953
Report Number1219602-2020-01907
Device Sequence Number2895698
Product Code MAI
UDI-Device Identifier00885556710319
UDI-Public00885556710319
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K180361
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/06/2022
Device Model Number72204984
Device Catalogue Number72204984
Device Lot Number50853840
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/02/2020
Supplement Date Received by Manufacturer07/22/2021
Initial Report FDA Received Date11/20/2020
Supplement Report FDA Received Date08/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-