• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 SITE~RITE 8 ULTRASOUND SYSTEM CONSOLE WITH CONNECTOR; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 SITE~RITE 8 ULTRASOUND SYSTEM CONSOLE WITH CONNECTOR; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770600
Device Problems Failure to Read Input Signal (1581); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/29/2020
Event Type  malfunction  
Manufacturer Narrative
The device has not been received by the manufacturer for evaluation.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per biomed, the unit is freezing and locking up and the buttons on the probe are not responsive.
 
Event Description
Per biomed, the unit is freezing and locking up and the buttons on the probe are not responsive.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of the unit is freezing and locking up and the buttons on the probe are not responsive was unconfirmed.During the initial phase of the assessment, the sr8 console was powered on for 24 hours.The freezing issue could not be reproduced.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.A history review of serial number (b)(6) showed no other similar product complaint(s) from this serial number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE~RITE 8 ULTRASOUND SYSTEM CONSOLE WITH CONNECTOR
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key10880165
MDR Text Key219372279
Report Number3006260740-2020-20623
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741138263
UDI-Public(01)00801741138263
Combination Product (y/n)N
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 04/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770600
Device Catalogue Number9770600
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2020
Date Manufacturer Received04/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-