MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION 4000 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION 4000 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938742
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that twenty-six (26) units of 4000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bags were leaking coming at the port.The leaks were discovered during setup and preparation.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
Additional information was added to d10, h3, h4, and h6.H10: only three (3) actual samples were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A functional testing was performed with tap water which revealed a leak between the spike port cap tube and the spike port bonding area of all 3 samples.The reported condition was verified for the returned samples.The cause of the condition was not determined; however, the most likely cause was due to inadequate or lack of cyclohexanone applied during the manufacturing process.The remaining twenty-three (23) samples were not received for evaluation; therefore, a device analysis could not be completed for those samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4000 ML TPN BAG
Common Device Name
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10880215
Report Number1416980-2020-07287
Device Sequence Number3803291
Product Code LHI
UDI-Device Identifier00085412477268
UDI-Public(01)00085412477268
Combination Product (Y/N)Y
Initial Reporter CountryCA
PMA/510(K) Number
K900585
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/31/2021
Device Catalogue NumberH938742
Device Lot Number60148501
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2020
Initial Date Received by Manufacturer 10/28/2020
Supplement Date Received by Manufacturer12/09/2020
Initial Report FDA Received Date11/20/2020
Supplement Report FDA Received Date12/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-