| Catalog Number |
H938742 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that twenty-six (26) units of 4000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bags were leaking coming at the port.The leaks were discovered during setup and preparation.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to d10, h3, h4, and h6.H10: only three (3) actual samples were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A functional testing was performed with tap water which revealed a leak between the spike port cap tube and the spike port bonding area of all 3 samples.The reported condition was verified for the returned samples.The cause of the condition was not determined; however, the most likely cause was due to inadequate or lack of cyclohexanone applied during the manufacturing process.The remaining twenty-three (23) samples were not received for evaluation; therefore, a device analysis could not be completed for those samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|