| Model Number |
VBH110502A |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/19/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional manufacturer narrative: cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium api, which is covalently bound to the device surface, and is essentially non-eluting.
|
| |
|
Event or Problem Description
|
|
The following information was reported to gore: on (b)(6) 2020, a patient was undergoing a procedure of unknown type with a gore® viabahn® endoprosthesis.After inserting the device, a separation between the stent and the catheter was noticed.Additionally, the stent was in a position where it would not deploy properly the device and delivery system were removed, and another device was utilized successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
The information initially reported stated the stent had separated from the delivery catheter.After the device was returned an engineering evaluation was performed.It was identified that the stent moved on the distal shaft but did not separate from the delivery catheter.Therefore this report will be retracted.The gore® viabahn® endoprosthesis instructions for use states, ¿do not withdraw the gore® viabahn® endoprosthesis back into the introducer sheath after the endoprosthesis is fully introduced¿.
|
| |
|