Model Number 3501430
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 03/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: material rupture.(b)(4).
 
Event or Problem Description
It was reported that a female patient, who underwent breast reconstruction with a 225cc mentor smooth round spectrum saline breast prosthesis on the left side, suffered left breast implant deflation, post-procedure.The deflation was diagnosed via visual examination.As a result, the patient underwent bilateral removal and bilateral replacement with two unspecified mentor saline breast prostheses on (b)(6) 2020.The patient was in stable condition after the surgery.
 
Additional Manufacturer Narrative
On december 15, 2020, the mentor failure analysis lab received the device for evaluation.On december 22, 2020, the product investigation was completed.Device investigation summary: during visual evaluation, an anomaly was observed on the device consistent with a crease/fold on the posterior view.A tear was also observed within the crease/fold measuring approximately 3.5 cm.The evaluation determined that the loss of shell integrity is consistent with a crease fold deflation of the implant shell, which is a known inherent risk of saline-filled mammary prosthesis.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Creating wrinkles or folds should be avoided during implantation or other procedures as it can result in a deflation in a later time.Breast implants may also simply wear out over time.A second product was received (lot-7671251).No adverse events were reported for this concomitant (contralateral) device, therefore no further analysis is required.As part of our quality process, the manufacturing records of this lot-serial number were reviewed and the manufacturing standards were met prior to the release of this lot.Each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SMOOTH ROUND SPECTRUM
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key10880368
Report Number1645337-2020-14928
Device Sequence Number2863686
Product Code FWM
UDI-Device Identifier00081317001973
UDI-Public00081317001973
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/20/2020
Device Model Number3501430
Device Catalogue Number3501430
Device Lot Number7342913
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2020
Initial Date Received by Manufacturer 10/29/2020
Supplement Date Received by Manufacturer12/15/2020
Initial Report FDA Received Date11/20/2020
Supplement Report FDA Received Date12/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age36 YR
-
-