| Model Number |
PM3140 |
| Medical Device Problem Code |
Pacing Problem (1439)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Based on the information available at this time, the specific complaint file covered by this investigation does not warrant capa escalation individually.If new information is received, the need for corrective action will be reassessed.Device complaints will be monitored through tracking and trending and escalated to capa when appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that an episode of high ventricular rate (hvr) was observed on the device.The device was reprogrammed to resolve the event.The patient was stable and there were no adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b3.
|
| |
|