| Model Number |
MC2-5090S |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/27/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned for evaluation as it remains in-situ.Root cause is unable to be determined at this time.
|
| |
|
Event or Problem Description
|
|
Information was received that x-ray images revealed that the end cap had disengaged.No revision procedure has been scheduled at this time.There was no patient injury reported.
|
| |
|
Event or Problem Description
|
|
See updated information in h2 and h10.
|
| |
|
Additional Manufacturer Narrative
|
|
No product has been returned for evaluation as it remains in-situ.Radiographs provided confirmed the alleged event.Even though no product has been returned, fsca investigation was performed and variation in the torque applied to the threaded cap during the assembly process was identified as the root cause of the failure.Per the manufacturing instructions, the threaded cap must be tightened to 40 in-lbs.Although all operators followed the assembly procedure, and the torque wrench indicated 40 in-lbs was applied, the manner of using the torque wrench resulted in variances in applied torque.The manner in which the operator handles the torque wrench may impact the actual torque applied to the end cap, thereby creating a false positive that the specified torque has been applied.If the specified torque is not applied, the effectiveness of the cap tightening process may be compromised.The information has been added to the database for tracking and trending purposes.
|
| |
|