| Model Number |
MMT-551NAL |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pump received with broken reservoir tube lip and cracked battery tube threads.A test reservoir was installed and did not lock in place due to complete broken reservoir tube lip.(b)(4).
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the retainer ring came off.Customer stated that the reservoir was able to lock in place when inserted into pump.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|