• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 0 DEG XLPE ACET LNR 40MM X 58MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 0 DEG XLPE ACET LNR 40MM X 58MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Model Number 71338680
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Osteolysis (2377); Impaired Healing (2378); Subluxation (4525); Joint Laxity (4526)
Date of Event 07/09/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Complaint reference: (b)(4).
 
Event or Problem Description
It was reported that the patient underwent revision surgery due to a loose, infected femoral component.During this procedure the implanted liner was removed and a antibiotic spacer was put in place.The patient then underwent revision surgery due to a spacer dislocation (5 or 6 weeks later due to timeframe of 2-stage revision).The whole s&n set being removed and replaced.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, the reported loosening of the patient¿s implant was likely due to the bone deficiency related to his previous infection.However, review of the provided documentation does not provide insight into the root cause of the reported infection, nor the dislocated antibiotic spacer.Without all supporting lab/pathology results, imaging, and/or the analysis of the explanted components, the source of the reported clinical reactions cannot be confirmed, and it cannot be concluded that they were associated with a mal-performance of the implant as the reported infections were years after implantation and therefore likely hematogenously spread.The patient impact beyond the revision procedure and related post-operative healing cannot be determined.No further medical assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.A review of the sterilization records revealed the batch was sterilized within normal parameters.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors and/or some potential probable causes that could contribute to the reported event have been identified as but not limited to abnormal motion over time, bone degeneration, fit/sizing, lack of ingrowth, lifetime of device, and traumatic injury.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 0 DEG XLPE ACET LNR 40MM X 58MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10884190
Report Number1020279-2020-06652
Device Sequence Number11524592
Product Code MBL
UDI-Device Identifier00885556113240
UDI-Public00885556113240
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K093363
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/18/2023
Device Model Number71338680
Device Catalogue Number71338680
Device Lot Number13JM13255
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/28/2020
Supplement Date Received by Manufacturer06/28/2021
Initial Report FDA Received Date11/23/2020
Supplement Report FDA Received Date06/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-