| Model Number |
45-500 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/26/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: analysis of the returned device identified: underweight, crease fold, red particles in the shell and opening on radius.A visual and microscopic analysis was performed which identified: crease sharp opening and stress marks.Based on the device analysis the final assessment is: a pattern of crease sharp on radius side of opening shell assessed as fold flaw opening.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|