| Model Number |
PED-450-25 |
| Medical Device Problem Codes |
Unintended Movement (3026); Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a pipeline failure to open.The patient was being treated for an unruptured saccular aneurysm of the left internal artery posterior communicating segment.The aneurysm max diameter was 15.6mm and the neck diameter was 6.8mm.Vessel tortuosity was severe.Dual antiplatelet treatment (dapt) was administered and pru level was 40.It was reported that the marksman was used to deliver a pipeline stent device.The distal tip would not open even with repeated attempts.The pipeline was accidentally deployed so it was withdrawn from the body.When removed from the catheter, it was noted that the tip of the pipeline was deformed.A new microcatheter and different stent were then used to complete the procedure.There was no harm or injury to the patient.
|
| |
|
Event or Problem Description
|
|
Additional information was received reporting that the first pipeline (model: ped-450-25, lot: b027924) was positioned in a vessel bend when it failed to open.The device was resheathed twice.Additional steps were taken to try and open the pipeline: pushed and pulled the guidewire, deployed at a longer distance, "retrieved it from the horizontal second of the brain", but all attempts failed.The marksman catheter jumped during retrieval causing the deployment.Continuous catheter flush was provided during the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
The distal and proximal dps restraints were found to be intact.The dps sleeves were found intact with no signs of damage.The distal hypotube and ptfe shrink tubing were found to be intact with no signs of elongation.The distal and proximal ends of the pipeline flex braid were fully opened and moderately frayed.No bend found on the pusher.No damages were found with the tip coil, distal marker, re-sheathing marker or with the proximal bumper.No other anomalies were observed.Based on the customer's photo, the customer complaint was confirmed to have failure to open at the distal end.However, based on the returned device, the pipeline flex was not confirmed to have "failure to open at the distal end" issue.The root cause could not be determined as the distal and proximal ends of pipeline flex braid were fully opened and moderately frayed.The damage to the braid on the ends of the pipeline flex braid is likely the results of the physician re-sheathing the device more than recommended two times.It is likely that the severe vessel tortuosity may have contributed to the failure to open issue.There was no non-conformance to specifications identified that led to the failure to open issue.Per our instructions for use (ifu): ¿begin to deliver the device using a combination of unsheathing the pipeline flex embolization device and pushing delivery wire simultaneously.After the distal end of the pipeline flex embolization device has successfully expanded, deploy the remainder of pipeline flex embolization device by pushing the delivery wire and/or unsheathing the pipeline flex embolization device.Resheathing and/or manipulation of the micro catheter, by locking down the delivery wire and moving both as a system, may facilitate expansion of the pipeline flex embolization device.The system is designed to allow for 2 full cycles of re-sheathing of the pipeline flex embolization device.Re-sheathing the pipeline flex embolization device more than 2 full cycles may cause damage to the distal or proximal ends of the braid.Do not use in patients in whom the angiography demonstrates the anatomy is not appropriate for endovascular treatment, due to conditions such as severe intracranial vessel tortuosity or stenosis.¿ medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|