| Model Number |
THX-165-028 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
10/27/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of the device labeling notes the following: the current overstitch¿ endoscopic suturing system (ess) instructions for use (ifu) addressed the known and potential event of "helix-could not be removed from tissue" as follows: warnings: only physicians possessing sufficient skill and experience in similar or the same techniques should perform endoscopic procedures.Troubleshooting: helix stuck in tissue: i.Use a suitable accessory through the primary channel to apply counter traction to the tissue around the helix, and pull the helix free.Ii.Once endoscopic techniques have been exhausted, utilize laparoscopic techniques to remove the helix.Adverse events: possible complications that may result from using the endoscopic suturing system include, but may not be limited to: conversion to laparoscopic or open procedure.Intra-abdominal (hollow or solid) visceral injury.
|
| |
|
Event or Problem Description
|
|
The tissue helix got stuck.It was ultimately unscrewed after a significant amount of effort and time.
|
| |
|