Model Number THX-165-028
Medical Device Problem Codes Entrapment of Device (1212); Difficult to Remove (1528)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 10/27/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the device labeling notes the following: the current overstitch¿ endoscopic suturing system (ess) instructions for use (ifu) addressed the known and potential event of "helix-could not be removed from tissue" as follows: warnings: only physicians possessing sufficient skill and experience in similar or the same techniques should perform endoscopic procedures.Troubleshooting: helix stuck in tissue: i.Use a suitable accessory through the primary channel to apply counter traction to the tissue around the helix, and pull the helix free.Ii.Once endoscopic techniques have been exhausted, utilize laparoscopic techniques to remove the helix.Adverse events: possible complications that may result from using the endoscopic suturing system include, but may not be limited to: conversion to laparoscopic or open procedure.Intra-abdominal (hollow or solid) visceral injury.
 
Event or Problem Description
The tissue helix got stuck.It was ultimately unscrewed after a significant amount of effort and time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVERSTITCH ENDOSCOPIC SUTURING SYSTEM
Common Device Name
SUTURING SYSTEM
Manufacturer (Section D)
APOLLO ENDOSURGERY
1120 s. capital of texas hwy
bldg 1, ste. 300
austin TX 78746
MDR Report Key10886663
Report Number3006722112-2020-00117
Device Sequence Number14177046
Product Code OCW
UDI-Device Identifier10811955020688
UDI-Public10811955020688
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K081853
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTHX-165-028
Device Catalogue NumberTHX-165-028
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/29/2020
Initial Report FDA Received Date11/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-