| Model Number |
09H39-01 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer reported a hemoglobin mismatch occurring on the cell-dyn emerald analyzer on an (b)(6) yr old male patient.Results provided: (b)(6) 2020 sid (b)(6) = 9.2 g/dl (reference range: 11-18 g/dl), another lab = 6.9 g/dl (reference range: 14-18 g/dl).All available patient information was included.Additional patient details are not available.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs and a review of labeling.No trends were identified for the customers issue.Labeling was reviewed and found to be adequate.Manufacturing documentation for the likely cause lot did not identify any issues associated with the complaint issue.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
|
| |
|