MAUDE Adverse Event Report: COOK ENDOSCOPY UTS PRECURVED ULTRA TAPER SPHINCTEROTOME; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY UTS PRECURVED ULTRA TAPER SPHINCTEROTOME; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number G22429
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/02/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter occupation: unknown.Pma/510(k) # k172665.Investigation evaluation: an evaluation of the device confirmed the report of incorrect cutting wire orientation.During our laboratory analysis, the sphincterotome was advanced through a duodenoscope that was placed in a simulated biliary position.The duodenoscope has an accessory channel that is 4.2 mm in diameter (olympus model number tjf-160v).The catheter exited the endoscope with the cutting wire facing 6 o¿clock.Prior to being bowed, the distal end entered the simulated papilla.The device was then attempted to be bowed and the cutting wire was facing 6 o'clock (appropriate orientation is approximately 11:00 - 1:00 o'clock).The device failed to bow.A visual inspection determined that there was a breakage at the center cannula, and a complete functional test could not be performed.The sphincterotome catheter was subjected to a close visual examination and twisting of the tubing was observed at the distal end.A discrepancy or anomaly that could have contributed to the reported event was not observed during our laboratory analysis of the returned product.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.A factor that can contribute to improper cutting wire orientation includes manipulating the handle with the catheter in a coiled position or with the precurved stylet inside the cannulating tip.The instructions for use advise the user: "upon removing device from package, uncoil and straighten sphincterotome." the user is then instructed to: "carefully remove precurved stylet from cannulating tip." the instructions for use contain the following comment: "note: do not exercise handle while device is coiled or precurved stylet is in place, as this may cause damage to sphincterotome and render it inoperable." prior to distribution, all uts precurved ultra taper sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During an endoscopic retrograde cholangiopancreatography (ercp) procedure, the physician used a cook uts precurved ultra taper sphincterotome.The orientation of the tip was bent in the opposite direction out of the package.The physician put it down the endoscope, but could not use it, pulled it back out of the endoscope, and used another to complete the procedure.No harm to the patient.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UTS PRECURVED ULTRA TAPER SPHINCTEROTOME
Common Device Name
KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key10886735
Report Number1037905-2020-00506
Device Sequence Number19540282
Product Code KNS
UDI-Device Identifier00827002224292
UDI-Public(01)00827002224292(17)230820(10)W4377626
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/20/2023
Device Model NumberG22429
Device Catalogue NumberUTS-20M
Device Lot NumberW4377626
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/02/2020
Initial Report FDA Received Date11/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-