Model Number SIG60CTAVM
Medical Device Problem Codes Failure to Form Staple (2579); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Date of Event 11/03/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a laparoscopic splenectomy, on the deposition of the spleen vessels, the artery and vein were removed with a magazine without traction, the instrument could be fired completely, there was poor staple formation, the staple line was incomplete centrally, after opening the reload, both vessel stumps were partially open as the staple line opened up and had to be overstitched.There was tissue damage.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the device had flush to sub flush, staple pushers on the proximal sections and the middle section of the cartridge due to an obstruction.It was reported that the staples did not form as expected, the deployed staples did not hold the tissue and the staple line opened up, and staples were missing from the staple line after firing.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur when an obstacle has been incorporated in the jaws during application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: ensure that no obstructions (such as clips) are incorporated in the instrument jaws.Firing over an obstruction may result in incomplete cutting action and/or improperly formed staples.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-STAPLE 2.0
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key10886746
Report Number1219930-2020-04945
Device Sequence Number19640495
Product Code GDW
UDI-Device Identifier10884521543690
UDI-Public10884521543690
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K160176
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Model NumberSIG60CTAVM
Device Catalogue NumberSIG60CTAVM
Device Lot NumberN9E0170Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2020
Initial Date Received by Manufacturer 11/04/2020
Supplement Date Received by Manufacturer02/09/2021
Initial Report FDA Received Date11/23/2020
Supplement Report FDA Received Date03/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-