• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TIBIAL SPACER BLOCK WIDE; PROSTHESIS,KNEE,PATELLOFEMOROTIBIAL,SEMI-CONSTRAINED,CEMENTED,POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TIBIAL SPACER BLOCK WIDE; PROSTHESIS,KNEE,PATELLOFEMOROTIBIAL,SEMI-CONSTRAINED,CEMENTED,POLY/METAL/POLY Back to Search Results
Model Number 74012646
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that in a tibial spacer block wide, the plastic ring inside is broken, therefore the orange shins do not click in.Tka procedure was performed.Procedure finished with the same device.This occurred after procedure.No surgical delay.No harm or injury reported on patient.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.Therefore, product analysis could not be performed at this time.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL SPACER BLOCK WIDE
Common Device Name
PROSTHESIS,KNEE,PATELLOFEMOROTIBIAL,SEMI-CONSTRAINED,CEMENTED,POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10886923
Report Number1020279-2020-06667
Device Sequence Number3006728
Product Code JWH
UDI-Device Identifier00885556003398
UDI-Public00885556003398
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K111711
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74012646
Device Catalogue Number74012646
Initial Date Received by Manufacturer 10/28/2020
Supplement Date Received by Manufacturer05/22/2021
Initial Report FDA Received Date11/23/2020
Supplement Report FDA Received Date05/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-