C.R. BARD, INC. (COVINGTON) -1018233 ULTRAFLEX SELF-ADHERING MALE EXTERNAL CATHETER Back to Search Results
Model Number 33302
Medical Device Problem Code Nonstandard Device (1420)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the patient having difficulty in removing the male external catheter.The patient already used a barrier film and a warm compress to remove.
 
Additional Manufacturer Narrative
The reported event was inconclusive.No sample was returned for evaluation.A potential root cause for this failure could be ¿plc failure¿.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "description/indication: the self-adhering male external catheter is designed for the management of male urinary incontinence.Contraindication: do not use on irritated or compromised skin.Precaution: do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply: 1) wash penis with mild soap and warm water.Dry thoroughly.2) trim pubic hair if necessary.3) unroll self-adhering catheter over penis.4) gently squeeze the catheter to properly seal adhesive to the skin.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.5) connect to drainage device.Directions: to remove: gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." the device was not returned.
 
Event or Problem Description
It was reported that the patient was having difficulty in removing male external catheter.Patient was already using a barrier film and a warm compress to remove.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX SELF-ADHERING MALE EXTERNAL CATHETER
Common Device Name
MALE EXTERNAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key10889707
Report Number1018233-2020-21152
Device Sequence Number3316142
Product Code NNX
UDI-Device Identifier00801741070754
UDI-Public(01)00801741070754
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/28/2025
Device Model Number33302
Device Catalogue Number33302
Device Lot NumberJUEU0650
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/30/2020
Supplement Date Received by Manufacturer01/05/2021
Initial Report FDA Received Date11/24/2020
Supplement Report FDA Received Date01/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-