MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG., BARREL BUR, 6 FLUTE F-SERIES - 5.0MM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG., BARREL BUR, 6 FLUTE F-SERIES - 5.0MM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number 0375951100
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the device broke during the procedure.The tip was retrieved.
 
Event or Problem Description
It was reported that the device broke during the procedure.The tip was retrieved.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: tip broke off probable root cause: inadequate window profile; inadequate welding process (i.E.Plasma, inductive, gluing); improper material strength; inadequate size selected for the application; material brittleness; excessive force applied by the user; incorrect rfid tag.The reported failure mode will be monitored for future reoccurrence.H3 other text: 81.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG., BARREL BUR, 6 FLUTE F-SERIES - 5.0MM
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key10891507
Report Number0002936485-2020-00487
Device Sequence Number19556916
Product Code GFA
UDI-Device Identifier37613327054850
UDI-Public37613327054850
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0375951100
Device Lot Number19170CE2
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/28/2020
Supplement Date Received by Manufacturer10/28/2020
Initial Report FDA Received Date11/24/2020
Supplement Report FDA Received Date01/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-