• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH LOCKING SCREW, T10 FULL THREAD, 2.7MM/L14MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH LOCKING SCREW, T10 FULL THREAD, 2.7MM/L14MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 657014S
Medical Device Problem Codes Positioning Failure (1158); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/29/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device discarded.
 
Event or Problem Description
As reported; "when orif procedure (right, foot joint) was performed as the initial operation, the head popped out from variax foot plate and did not lock.".
 
Event or Problem Description
As reported; "when orif procedure (right, foot joint) was performed as the initial operation, the head popped out from variax foot plate and did not lock.".
 
Additional Manufacturer Narrative
The reported event could not be confirmed.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A potential non-conformity report was initiated to address this event.The comprehensive root cause analysis of the potential ncr included a surface and microstructure analysis as well as nano-indent hardness measurement of variax1 and variax2 screw samples by the fraunhofer institute on behalf of stryker.Further internal investigation efforts were focusing on the anodization color difference between variax1 and variax2 screws.Significant differences between the screws which could result in the reported complaints have not been revealed in either fraunhofer or stryker investigations.Excessive in-house handling and bench testing with variax2 screws from various manufacturing batches showed that locking is achieved as intended.The desired increase in torque indicating to the customer that the screw is locking in the plate, can be confirmed.A review of the device history for the reported lot did not indicate any abnormalities.No deviation from the specifications was noted.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Considering the information given and based on the above investigations a root cause of the reported event could not be determined.If any further substantial information is provided, the investigation report will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW, T10 FULL THREAD, 2.7MM/L14MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
MDR Report Key10891535
Report Number0008031020-2020-02507
Device Sequence Number11536988
Product Code HWC
UDI-Device Identifier07613327071276
UDI-Public07613327071276
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K180500
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/25/2023
Device Model Number657014S
Device Catalogue Number657014S
Device Lot Number1000250542
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/29/2020
Supplement Date Received by Manufacturer03/04/2021
Initial Report FDA Received Date11/24/2020
Supplement Report FDA Received Date03/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-