| Model Number |
1825-1036S |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Nonunion/Delayed-union Bone Fracture (2369)
|
| Date of Event |
10/22/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Upon completion of investigation, additional information will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
The customer reported a broken nail.Patient arrives at the clinic with a broken nail and a broken distal screw.No trauma apparent, no further histological findings, non-union, no calcification visible in the fracture gap.Revision surgery with the "sliding nail" (company intercus) slide 12mm.
|
| |
|
Event or Problem Description
|
|
The customer reported a broken nail.Patient arrives at the clinic with a broken nail and a broken distal screw.No trauma apparent, no further histological findings, non-union, no calcification visible in the fracture gap.Revision surgery with the "sliding nail" (company intercus) slide 12mm.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed, since the product was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the received nail is completely broken from the center region.The fracture pattern resembles a fatigue fracture (evident by appearance of shiny surface).This just weakened the nail.The fracture pattern also suggests that after a while the breakage was abrupt and instantaneous with material chip off from the side.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on investigation, the root cause was attributed to a patient factor.The failure was caused by patient factor (such as bone quality, activity, non-union or high load application) resulted in fatigue fracture, which can be correlated by the communication of non-union, evidently can be confirmed based on implant & explant dates.¿these devices can break when subjected to the increased loading associated with delayed unions and/or non-unions.Internal fixation devices are load sharing devices which are intended to hold fractured bone surfaces in apposition to facilitate healing.If healing is delayed or does not occur, the appliance may eventually break due to metal fatigue.Loads on the device produced by load bearing and the patient¿s activity level will dictate the longevity of the device.¿ if any further information is provided, the complaint report will be updated.
|
| |
|