Model Number PS210-030S
Medical Device Problem Codes Fire (1245); Melted (1385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/29/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare professional (hcp) regarding a generator and handpiece.It was reported that the handpiece caught fire and the tip melted while the handpiece was being used during a case.There was no impact on patient outcome.
 
Additional Manufacturer Narrative
Analysis summary: complaint confirmed: upon receiving the device, the heat shrink was torn on one side and was shifted above the finger grip component more than it should have been.The heat shrink component is assembled onto the shaft of the blade component to protect the device and user.The device was observed and still tested for functionality to attempt to duplicate the complaint in the complaint lab and was unsuccessful with the device ¿catching on fire¿ and the heat shrink ¿melting¿ observing the heat shrink component, it shows no signs of melting but the device did malfunction due to the component not being properly assembled during manufacturing which contributed to why the device did not function as its intended purpose and use.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLASMABLADE 3.0S
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MEDTRONIC ADVANCED ENERGY (SALIENT)
180 international drive
portsmouth NH 03801
MDR Report Key10893112
Report Number1226420-2020-00148
Device Sequence Number14176919
Product Code GEI
UDI-Device Identifier00643169936683
UDI-Public00643169936683
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K093695
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2023
Device Model NumberPS210-030S
Device Catalogue NumberPS210-030S
Device Lot Number2002189
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/29/2020
Supplement Date Received by Manufacturer12/02/2020
Initial Report FDA Received Date11/24/2020
Supplement Report FDA Received Date12/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-