Analysis summary: complaint confirmed: upon receiving the device, the heat shrink was torn on one side and was shifted above the finger grip component more than it should have been.The heat shrink component is assembled onto the shaft of the blade component to protect the device and user.The device was observed and still tested for functionality to attempt to duplicate the complaint in the complaint lab and was unsuccessful with the device ¿catching on fire¿ and the heat shrink ¿melting¿ observing the heat shrink component, it shows no signs of melting but the device did malfunction due to the component not being properly assembled during manufacturing which contributed to why the device did not function as its intended purpose and use.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|