| Model Number |
401763 |
| Medical Device Problem Codes |
Failure to Capture (1081); Loss of Threshold (1633)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results, method and conclusion codes, along with investigation results will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
During tavr procedure, while rc pacing, there was loss of capture resulting in a new valve having to be redeployed.Troubleshooting included replacing the remington cables, but the issue persisted.The device was replaced, and the issue was resolved.Due to this event a new valve had to be redeployed.
|
| |
|
Additional Manufacturer Narrative
|
|
One 5f pacel bipolar pacing catheter (flow direct) was received for analysis.The results of the investigation concluded that there were no shorted, crossed, or intermittently open circuits.The measured resistance values for both the tip and ring electrode circuits were within specification.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.
|
| |
|