| Model Number |
N/A |
| Medical Device Problem Code |
Biocompatibility (2886)
|
| Health Effect - Clinical Codes |
Pain (1994); Swelling (2091); Reaction (2414); Limited Mobility Of The Implanted Joint (2671)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the decice remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Additional associated products and mdrs.00549003410 trabecular metal distal femoral augment block lot# 63322074.Mdr- 3005751028-2020-00099.00545001730 trabecular metal femoral diaphyseal cone, 30mm lot# 63392061.Mdr- 3005751028-2020-00101.00544705336 nexgen lcck tm coupled tibial cone, a/p wedged size 3 lot# 62380724.Mdr- 3005751028-2020-00102.
|
| |
|
Event or Problem Description
|
|
It was reported she has been encountering pain, swelling, inflammation, and difficulty walking and exercising since her revision surgery.She also stated she had a metal allergy test done, which determined she is highly allergic to nickel and cobalt chromium.
|
| |
|
Event or Problem Description
|
|
This follow-up report is being submitted to relay additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected.No product was returned; visual and dimensional evaluations could not be performed.Review of the device history record(s) for item/lot [00549003410/63141884] identified no deviations or anomalies during manufacturing.Reported event is not related to a combination of products; therefore, a compatibility review is not applicable.Review of complaint history identified 1 additional similar complaints for the reported item (00549003410) - the additional product on this same complaint- and no additional complaints for the reported part and lot combination (00549003410/63141884).Complaints are monitored through monthly complaint review (reference os 5017) in order to identify potential adverse trends.Medical records were not provided.A field action search using cognos bi identified no quality hold with an r-code as the reason associated with the following part/lot combinations (00549003410/63141884) as of 7-jan-2020.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.
|
| |
|