| Model Number |
DADE ACTIN FSL ACTIVATED PTT REAGENT |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/26/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer contacted a siemens customer care center (ccc).Quality controls (qc) recovered in range on both days, and the base height (bh) and the delta height (dh) for the qc compared well.However, significant differences were observed when comparing the bh and dh levels for the patient sample aptt results from (b)(6) 2020 and (b)(6) 2020.This indicates that a potential sample handling or pre-analytical issue may have occurred on (b)(6) 2020.The cause of the event is unknown.The reagent is performing according to specification.No further evaluation of this device is required.
|
| |
|
Event or Problem Description
|
|
Discordant, falsely elevated activated partial thromboplastin time (aptt) results were obtained on six patient samples on a sysmex ca-660 system using dade actin fsl activated ptt reagent.The discordant results were reported to the physician(s).The samples were refrigerated and rerun the following day for aptt, as the customer was questioning the results.The repeated samples recovered lower.These repeat results were considered to be the correct results.There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated activated partial thromboplastin time (aptt) results.
|
| |
|