| Model Number |
220 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Vascular Dissection (3160)
|
| Date of Event |
09/23/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age at time of event - 18 years or older.
|
| |
|
Event or Problem Description
|
|
It was reported that the device locked on the wire and a dissection occurred.A 5f x 11cm super sheath introducer sheath with 0.038in mini guidewire was selected for use in a procedure within the femoral artery.The post-dilated target lesion was 300mm long with no calcification.During the procedure, the mini guidewire locked on to the non-boston scientific introducer needle.The super sheath introducer sheath was never advanced inside the body; only the mini guidewire that was provided with the device was inside the body.There was bleeding from a small rift in the vessel that required extra hand compression.The procedure was completed with another of the same device.No other patient complications were reported.
|
| |
|