DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Gas Output Problem (1266)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/02/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).A getinge field service engineer (fse) evaluated the iabp, and was unable to reproduce the reported issue.However, in reviewing the alarm logs, the fse verified 2 "auto fill failure", and 2 "gas gain in iab circuit." the fse determined that the helium tank was low and replaced it with a customer issued part.The fse concluded that low helium in the tank could likely contribute to an autofill failure as a result of insufficient inflation in the balloon, and thereby causing a gas gain/loss in the circuit.The fse performed a function test of the unit without any issues.The fse then successfully performed a full preventative maintenance (pm) service, and performed all functional and safety checks to meet factory specifications.The iabp was then released to the customer for return to clinical service.
|
|
Event Description
|
It was reported that prior to use on a patient, the cardiosave intra-aortic balloon pump (iabp) had a "gain error" issue.However, there was no patient involvement, and no adverse event reported.
|
|
Event Description
|
It was reported that prior to use on a patient, the cardiosave intra-aortic balloon pump (iabp) had a "gain error" issue.However, there was no patient involvement, and no adverse event reported.
|
|