Model Number CV-180
Device Problem No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 07/09/2020
Event Type  malfunction  
Manufacturer Narrative
The device was evaluated by olympus and found the patient circuit board failure caused no image.The reason for the circuit board failure is unknown.This device is an olympus asset, it was repaired and returned to asset inventory.An olympus (b)(4) capa has been opened to manage the actions related to remediation of this late mdr reporting.
 
Event Description
The device was returned to the olympus repair center for general inspection.During evaluation, the repair center identified the device presented no image due to a printed circuit board failure.There was no patient involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, since more than 14 years have passed since manufacturing, it is likely the parts on the board have deteriorated over time due to repeated use for a long period of time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10900342
MDR Text Key218224473
Report Number8010047-2020-09420
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170215513
UDI-Public04953170215513
Combination Product (y/n)N
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/09/2020
Initial Date FDA Received11/25/2020
Supplement Dates Manufacturer Received01/05/2021
Supplement Dates FDA Received01/14/2021
Is This a Reprocessed and Reused Single-Use Device? No