Model Number CLH-SC |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Omsc confirmed that more than 15 years had passed since the device was manufactured.Therefore, omsc guessed that the reported event was caused by aging.If additional information becomes available, this report will be supplemented.
|
|
Event Description
|
Olympus inspected the device at the olympus service operation repair center and found that the ac inlet of the device was scorched.There was no report of patient injury associated with the event.
|
|
Manufacturer Narrative
|
This mdr is being submitted as part of a retrospective review and remediation effort based on enhancements made to the company¿s mdr and complaint handling processes.Capas have been opened to manage the actions that are being taken to remediate this issue and ensure any required mdr reporting is completed.Correction to g3 of the initial medwatch.The aware date should be 21-jul-2020.Olympus will continue to monitor the field performance of this device.
|
|